The autoclave display tells you that the cycle has finished – but not whether steam actually reached every instrument. Sterilisation monitoring tests confirm this, and they also form the basis of your records in case of an official inspection.

Why the display is not enough

The display measures chamber temperature and pressure. Inside pouches, in hollow instruments or on overloaded trays, however, air can remain trapped, and there steam cannot reach the surface. Monitoring tests check steam penetration and the actual sterilisation conditions.

Bowie-Dick test: every day, before the first cycle

In pre-vacuum (class B) autoclaves, the Bowie-Dick test checks whether the unit removes air properly and whether steam penetrates evenly. It is usually run at the start of the working day, with an empty chamber, in a dedicated test cycle. If the colour change on the test sheet is uneven, the autoclave must not be used until the cause has been found.

Helix test for hollow instruments

Steam reaches the inside of handpieces, suction tubes and other narrow-lumen devices less easily. The helix test (a process challenge device, PCD) simulates this difficult situation with an indicator placed at the end of a long, thin tube, showing whether the cycle reliably sterilises hollow instruments too.

Chemical indicators: in every pack

Chemical indicators show sterilisation conditions by changing colour. The process indicator printed on pouches only shows that the pack has been through the autoclave; more advanced (integrating and emulating) indicators also show whether time, temperature and steam together reached the required values. For critical instruments, it is worth placing an indicator inside the pack as well.

Biological indicators: periodic verification

A biological indicator contains highly resistant bacterial spores (Geobacillus stearothermophilus) and proves that the cycle actually kills microorganisms. How often to use it depends on the manufacturer, your facility’s hygiene protocol and the national regulations in force; it is always advisable after a repair or service.

Seal check

A successful cycle is of little use if the pouch seal is faulty. A seal-check test verifies the quality of the seam produced by your sealing machine – typically daily or as specified by the manufacturer.

Documentation

Quick checklist

You can order sterilisation monitoring tests and sterilisation pouches at net prices with a VAT invoice in the Neximed webshop: Sterilisation monitoring tests and Sterilisation pouches and reels. For instrument preparation, see our earlier article: Disinfecting and sterilising instruments in the practice.