Dental handpieces, contra-angles and turbines are among the most expensive instruments in a practice and are used every day. Their lifespan depends largely on whether the team consistently cleans, lubricates and sterilises them after each treatment. Regular maintenance not only reduces repair costs but also supports patient safety.

Immediately after treatment

At the end of treatment, before removing the bur or head, run the handpiece with water and air for a few seconds to flush debris out of the internal channels. Then wipe the outer surface with an alcohol-based disinfectant wipe approved by the manufacturer. Never immerse a handpiece in disinfectant solution unless the manufacturer explicitly allows it.

Internal cleaning and lubrication

Before sterilisation, the handpiece must also be cleaned and lubricated internally, using a dedicated handpiece oil spray or an automatic maintenance unit. Apply the oil through the connector specified by the manufacturer, then remove the excess by briefly running or blowing out the handpiece. Too little oil accelerates bearing wear, while too much can bake on during sterilisation.

Sterilisation

Common mistakes

The most frequent problems are skipped lubrication, incorrect air pressure, running without a bur and using damaged burs. If a handpiece becomes noisier, heats up, vibrates or the bur no longer locks securely, stop using it and send it for service. Having several handpieces per treatment unit leaves enough time for a complete maintenance cycle.